A hemacytometer (also spelled hemocytometer) is a laboratory instrument used to count cells for purposes of determining their concentration in a liquid mixture. The instrument's name is derived from blood cells; however, it can be used to count any cell of similar size.
Principles
The hemacytometer is designed to be used as a (particularly thick) glass microscope slide. Its primary feature is a rectangular indentation that creates a chamber of certain dimensions. This chamber is etched with a grid of perpendicular lines.
When a liquid sample containing immobilized cells is placed on the chamber, it is covered with a cover glass, and capillary action completely fills the chamber with the sample. Looking at the chamber through a microscope, the researcher can count the number of cells in the chamber. Different kinds of cells can be counted separately as long as they are visually distinguishable. The number of cells in the chamber is used to calculate the concentration of the cells in the mixture from which the sample was taken: it is the number of cells in the chamber divided by the chamber's volume (the chamber's volume is known from the start), taking account of any dilutions and counting shortcuts:
concentration of cells in original mixture =
Usage tricks
- Mix your original mixture thoroughly before taking a sample. This ensures that your sample is representative, and not just an artifact of the particular region of your original mixture from which you drew it.
- Use an appropriate dilution of your mixture with regard to the number of cells you hope to count. If your sample is not diluted enough, the cells will be too crowded and difficult to count. If it is too dilute, your sample size will not be enough to make strong inferences about the concentration in the original mixture. Naturally, you must have a rough idea of the concentration before you begin in order to guess an appropriate dilution. If your mixture is colored, it may be helpful to memorize a particular intensity of that color at which the mixture tends to be easy to analyze.
- Analyze multiple chambers. By performing a redundant test on a second chamber, you can compare the results. If they differ greatly, your method of taking the sample may be unreliable (e.g. the original mixture is not mixed thoroughly). You can use the average of your results for a more accuate calculation.
- Make sure to put enough liquid on the instrument that some leaks out of the cover glass when it is placed over the chamber. Otherwise, it is uncertain whether the space under the cover glass is completely filled with liquid. This volume should be the same every time you use the instrument.
- Do not use a paper wipe to dry the excess liquid. The same capillary action that filled the chamber will then dry it out.
- Watch out for the objective lens. Remember that the hemacytometer is thicker than a normal microscope slide. If you focus too closely, your objective lens may contact the instrument. This may affect your choice of which objective lens you use - carefully figure out what will fit, before you start.
- Count across the rows or down the columns. Use the gridlines to help you remember which areas' cells have already been counted.
- You don't have to count the whole chamber. If there are lot of cells, you can just perform your count in a section of the chamber and use the grid to determine what proportion of the chamber that is. You can then extrapolate to estimate how many cells are in the chamber, and use that figure in your final calculation. This gives you speed at the expense of potential accuracy; if possible, using a more appropriate dilution is better.
- Are the lines in or out? Some cells inevitably fall smack on top of the outside gridlines that mark the edges of the chamber. If you plan a standard routine that doesn't favor either option (e.g. cells on the top and bottom lines can be counted, but not cells on the left or right lines) before you start and abide by it thereafter, you are not adding any bias to your results.